Carbon dioxide monitor

QR code
share
Brand Landswick
Detail

1 (1).jpg1 (2).jpg


Product Features and Advantages

· It only takes 5 seconds from start-up to monitoring to the first end-of-call data, necessary for first aid and CPR

· Compact, lightweight, one-handed operation

· In the large character mode, the end-call data is clearly visible at a distance of 5 meters

· It can be measured after starting up without zero calibration; the atmospheric pressure is automatically compensated, and it can provide accurate monitoring regardless of plateau or helicopter rescue

· Multiple data output: USB, Bluetooth (optional); multiple data processing and recording modes

· Optional interference gas compensation, lung pressure and body temperature compensation to meet higher precision requirements



Applicable environment

Pre-hospital monitoring, transfer monitoring, first aid and CPR monitoring, emergency department, operating room, intensive care, neonatal monitoring, sleep monitoring, etc.

Charge code: 310601009 Charge name: Carbon dioxide response curve 20-50 yuan/item (Class A medical insurance)



Clinical significance of carbon dioxide monitoring

Indications for monitoring: safe application of anesthesia machines and ventilators; various types of respiratory insufficiency; cardiopulmonary resuscitation; severe shock; heart failure and pulmonary infarction;



Technical Parameters

PrincipleNDIR(Non-dispersive infrared technology)
Power supplyAC input: 220V, 50Hz AC power consumption: not more than 10VA
Working conditionsTemperature: 5℃~40℃, Humidity: 30%~75%, Atmospheric pressure: 86kPa-106kPa
BatteryLithium battery: 3.7V 1.35AH, fully charged continuous working time: 10 hours (approx.), charging time: no more than 4 hours
Size and weightDimensions: 55mm(W)x44mm(H)x44mm(D), Weight: 80g
Show1.5' color TFT
Carbon dioxide concentration

Measuring range: 0-150mmHg (0-20kPa, or 0-20% (V/V))

Accuracy: ±2mmHg when 0-40mmHg, ±5%(reading)when>41-70mmHg , ±8%(reading)when>71-100mmHg, ±10%(reading)when>101-150mmHg

EtCO₂ update: per breath, trend graph length: 24 hours

Respiratory rateMeasuring range: 3-150 t/min, accuracy: ±1% of reading or ±1 times/min whichever is greater, trend graph length: 24 hours
Audible alarm

End-tidal carbon dioxide and respiratory rate exceed the limit alarm, and the upper and lower limits of the alarm are adjustable. Suffocation alarm, power shortage and abnormal technical status alarm , alarm volume can be adjusted in 8 levels